Analytical methods and scope
The instruments, columns, gradients, reference materials and reporting limits behind every number we print. Published in full so a technical buyer can evaluate the work rather than trust the brand.
Last updated September 2026
Accreditation status. The laboratory is built for ISO/IEC 17025 and is not yet accredited; it will apply once it is operating. Until the certificate issues, no result on this site or on any certificate we produce is an accredited result, and we do not describe it as one. The accredited scope and certificate number will be published on this page on issuance.
Method detail
Each method below lists what it actually runs on. Where a compendial standard governs the method we name the chapter. Where no compendial standard exists for a research peptide, we say that plainly instead of implying one.
Purity by reversed-phase HPLC
USP <621> chromatography; area normalization
HPLC-01, quaternary pump with diode-array detector
| Column | C18, 4.6 × 250 mm, 5 µm, 300 Å pore |
| Mobile phase A | 0.1% trifluoroacetic acid in water |
| Mobile phase B | 0.1% trifluoroacetic acid in acetonitrile |
| Gradient | 5% to 65% B over 30 min, then 95% B wash and re-equilibration |
| Flow rate | 1.0 mL/min |
| Column temperature | 30 °C |
| Detection | 214 nm (peptide bond), 280 nm secondary channel |
| Injection volume | 20 µL at nominal 1.0 mg/mL |
| Replicates | Triplicate injections; mean reported with RSD, acceptance RSD ≤ 2.0% |
How it is reported. Reported as area percent of total peptide-related peaks at 214 nm, as the mean of three injections with the relative standard deviation printed alongside it. A purity figure from a single injection cannot be distinguished from a bad injection, so we do not report one. Peaks below a 0.05% reporting threshold are excluded, as are non-peptide peaks including counter-ion and solvent fronts.
Identity confirmation by LC-MS/MS
Mass confirmation against theoretical monoisotopic and average mass
LCMS-01, electrospray ionization, positive mode
| Ionization | ESI positive, capillary 3.5 kV |
| Scan range | 300 to 2000 m/z |
| Charge states assessed | [M+2H]²⁺ and [M+3H]³⁺, minimum two states |
| Acceptance | Observed within 0.1 Da of theoretical for each assessed charge state |
| Deconvolution | Average neutral mass reported alongside observed m/z |
How it is reported. Identity is confirmed by mass measurement, not by retention-time comparison alone. Both observed and theoretical values are printed on the certificate so the confirmation can be checked by the reader.
Net peptide content
External calibration against a qualified reference standard
HPLC-01 with quantitative UV detection
| Calibration | Five-point external curve, bracketed |
| Linearity acceptance | r² ≥ 0.995 across the working range |
| Reference material | USP reference standard where one exists, otherwise a characterized peptide standard with certificate |
| Replicates | Duplicate preparations, mean reported, RSD ≤ 2.0% |
How it is reported. Reported in milligrams per vial and as a percentage of label claim. Net peptide content excludes water, counter-ion and residual solvent, so it is routinely lower than a gross mass measurement.
Bacterial endotoxin
USP <85> Bacterial Endotoxins Test
Kinetic chromogenic LAL, cartridge-based reader (Endosafe nexgen-PTS)
| Method | Kinetic chromogenic |
| Lysate sensitivity | 0.005 EU/mL |
| Interference control | Positive product control recovery 50% to 200% required |
| Dilution | Sample diluted to a non-interfering concentration below the MVD |
How it is reported. Reported in EU/mg of peptide. No universal endotoxin threshold exists for research-use materials, so the value is reported for the submitter's own assessment rather than judged against a limit we invented.
Elemental impurities
USP <232> limits, USP <233> procedure
ICP-MS with collision cell
| Elements | Cadmium, lead, arsenic, mercury (Class 1) |
| Preparation | Closed-vessel microwave digestion in nitric acid |
| Internal standards | Scandium, rhodium, iridium |
| Quantitation limit | 0.05 µg/g per element |
| Spike recovery acceptance | 70% to 150% |
How it is reported. Reported in µg/g against the USP <232> parenteral option where the submitter designates one. Otherwise reported as measured concentrations.
Microbial screen
USP <61> and <62> enumeration and specified organisms, with qPCR bioburden screen
Real-time PCR thermocycler (QuantStudio class)
| Total aerobic microbial count | USP <61>, plate count |
| Total yeast and mould count | USP <61>, plate count |
| Specified organisms | USP <62>, absence testing |
| Rapid screen | qPCR bioburden panel for a same-week indication |
How it is reported. This is a microbial screen. It is not a USP <71> compendial sterility test, which requires a 14-day incubation and a defined aseptic sampling plan. Any certificate carrying this method says so in the method column. If you need USP <71>, ask for it by name and we will quote it as referral testing under ISO/IEC 17025 clause 6.6.
Fentanyl and analog screen
Targeted LC-MS/MS screen (MRM) with confirmation on the same run
LCMS-01, targeted MRM screen with confirmation on the same run
| Screen cutoff | 50 ng/mL |
| Confirmation | A positive requires qualifier ion ratios and retention time to match a reference standard |
| Reporting convention | ND indicates not detected at the stated cutoff |
How it is reported. A not-detected result means the analyte was below the stated cutoff. It is not a statement that the substance is absent at any concentration, and the certificate says so.
Sterility: rapid screen and referred compendial test
qPCR microbial DNA screen in house; USP <71> by an accredited partner laboratory
Real-time PCR thermocycler (QuantStudio class)
| Rapid screen | Broad-range bacterial and fungal DNA targets, reported as Detected or Not detected |
| Compendial test | USP <71> membrane filtration or direct inoculation, 14 days, by the partner laboratory |
| On the certificate | The screen is named as a qPCR screen. A referred USP <71> result names the performing laboratory, its accreditation and its report number |
How it is reported. A qPCR screen detects DNA and observes no growth, so it is never reported as "no growth" or as a USP <71> result. When USP <71> is ordered, an interim certificate issues with sterility pending and the final certificate replaces it under the same accession.
Residual trifluoroacetic acid
Ion chromatography with suppressed conductivity
Ion chromatograph with conductivity detector
| Reported as | % w/w of the lyophilised solid |
| Limits | LOD and LOQ printed on every result |
How it is reported. TFA is a counter-ion left by reverse-phase purification. It is reported, not judged, unless the submitter supplies a specification.
Residual solvents
Headspace GC-MS against ICH Q3C class limits
Headspace sampler with single-quadrupole GC-MS
| Analytes | Common purification and synthesis solvents, each with its own LOQ |
| Specification | ICH Q3C permitted daily exposure limits, converted to ppm |
How it is reported. Each analyte is listed with its limit of quantitation. ND means below that limit.
Container closure integrity and solubility
Dye ingress following the approach of USP <1207>; visual solubility using USP General Notices 5.30 terms
Vacuum chamber, dye bath, visual inspection station
| Closure integrity | Per vial, reported as integral or not integral |
| Solubility | Reconstitution in water at the labelled volume, described in compendial terms |
How it is reported. Dye ingress is a probabilistic method: a pass means no ingress was seen under the stated challenge, not that a leak of any size is impossible.
Unit-to-unit variance
Net peptide content on additional vials, each prepared and assayed separately
HPLC-01 against a qualified reference standard
| Vials | Three vials beyond the one used for the main assay |
| Reported as | Each vial's value, then the mean and the RSD across vials |
How it is reported. The RSD across vials includes both fill variation and method variation. The method's own repeatability is established during validation, so the two can be told apart.
Stability program
Informed by ICH Q1A(R2) storage and timepoint design
Controlled chambers, analysis on HPLC-01 and LCMS-01
| Storage conditions | 2 to 8 °C refrigerated, and 25 °C / 60% RH ambient challenge |
| Timepoints | 30, 60 and 90 days from accession |
| Assays at each point | Purity and net peptide content, against the day-zero result |
| Retained material | Drawn from the same retained aliquot as the original test |
How it is reported. Each timepoint issues its own certificate referencing the day-zero accession, so the degradation curve is a chain of verifiable documents rather than a single summary claim.
Replicates
Every quantitative result on our certificates is the mean of replicate runs, printed with the relative standard deviation so you can see the spread rather than take the average on trust.
This matters more than it sounds. A single injection produces one number and no way to know whether that number is real. A bad injection volume, an air bubble, carryover from the previous sample or noisy peak integration all produce a plausible-looking figure. Three injections reading 97.4, 97.6 and 97.5 tell you the method is under control. Three reading 97.4, 94.1 and 98.0 tell you to investigate rather than to issue a certificate.
| Basis | What varies between runs | What it demonstrates |
|---|---|---|
| Replicate injections | The same prepared vial is injected more than once. | Instrument and integration precision. Used for purity. |
| Replicate preparations | The sample is weighed and dissolved separately for each run. | Precision of the whole method, including preparation. Used for net peptide content. |
The certificate states which basis applies to each result and prints n and the RSD next to the value. Where an RSD exceeds the method acceptance criterion, the run is investigated and repeated. It is not averaged into something presentable.
Sampling and what a result applies to
Under ISO/IEC 17025 clause 7.3, responsibility for sampling sits with whoever performed the sampling. Samples are selected and submitted by the client, so every result applies to the unit as received. It is not evidence about any other unit of that lot, and we do not present it as such. That limitation is printed on the certificate rather than left for the reader to infer.
Where a shipment arrives in a condition that could affect the result, the deviation is recorded and the certificate states it. See sample handling and retention for the condition assessment we run on receipt.
Retained material and re-testing
Every accessioned sample is aliquoted and retained for 90 days under controlled storage. Any client may request a re-test on the retained aliquot within that window, which is the only way to settle a disagreement about a result without introducing a new sample and a new set of variables. The full policy is in sample handling and retention.
Quality control on every run
A known reference standard is analysed alongside client samples on each analytical run, and its recovery is recorded. We will publish that record, with a chart of our own accuracy over time and turnaround times computed from the same database that issues the certificates, once the laboratory is operating.
Referral testing
Work outside our scope, including USP <71> compendial sterility, is placed with a subcontracted laboratory under ISO/IEC 17025 clause 6.6. Referred work is identified as referred on the certificate, and the performing laboratory is named. We do not present another laboratory's work as our own.