Analytical methods and scope

The instruments, columns, gradients, reference materials and reporting limits behind every number we print. Published in full so a technical buyer can evaluate the work rather than trust the brand.

Last updated September 2026

Accreditation status. The laboratory is built for ISO/IEC 17025 and is not yet accredited; it will apply once it is operating. Until the certificate issues, no result on this site or on any certificate we produce is an accredited result, and we do not describe it as one. The accredited scope and certificate number will be published on this page on issuance.

Method detail

Each method below lists what it actually runs on. Where a compendial standard governs the method we name the chapter. Where no compendial standard exists for a research peptide, we say that plainly instead of implying one.

PUR

Purity by reversed-phase HPLC

USP <621> chromatography; area normalization

HPLC-01, quaternary pump with diode-array detector

ColumnC18, 4.6 × 250 mm, 5 µm, 300 Å pore
Mobile phase A0.1% trifluoroacetic acid in water
Mobile phase B0.1% trifluoroacetic acid in acetonitrile
Gradient5% to 65% B over 30 min, then 95% B wash and re-equilibration
Flow rate1.0 mL/min
Column temperature30 °C
Detection214 nm (peptide bond), 280 nm secondary channel
Injection volume20 µL at nominal 1.0 mg/mL
ReplicatesTriplicate injections; mean reported with RSD, acceptance RSD ≤ 2.0%

How it is reported. Reported as area percent of total peptide-related peaks at 214 nm, as the mean of three injections with the relative standard deviation printed alongside it. A purity figure from a single injection cannot be distinguished from a bad injection, so we do not report one. Peaks below a 0.05% reporting threshold are excluded, as are non-peptide peaks including counter-ion and solvent fronts.

ID

Identity confirmation by LC-MS/MS

Mass confirmation against theoretical monoisotopic and average mass

LCMS-01, electrospray ionization, positive mode

IonizationESI positive, capillary 3.5 kV
Scan range300 to 2000 m/z
Charge states assessed[M+2H]²⁺ and [M+3H]³⁺, minimum two states
AcceptanceObserved within 0.1 Da of theoretical for each assessed charge state
DeconvolutionAverage neutral mass reported alongside observed m/z

How it is reported. Identity is confirmed by mass measurement, not by retention-time comparison alone. Both observed and theoretical values are printed on the certificate so the confirmation can be checked by the reader.

CONT

Net peptide content

External calibration against a qualified reference standard

HPLC-01 with quantitative UV detection

CalibrationFive-point external curve, bracketed
Linearity acceptancer² ≥ 0.995 across the working range
Reference materialUSP reference standard where one exists, otherwise a characterized peptide standard with certificate
ReplicatesDuplicate preparations, mean reported, RSD ≤ 2.0%

How it is reported. Reported in milligrams per vial and as a percentage of label claim. Net peptide content excludes water, counter-ion and residual solvent, so it is routinely lower than a gross mass measurement.

ENDO

Bacterial endotoxin

USP <85> Bacterial Endotoxins Test

Kinetic chromogenic LAL, cartridge-based reader (Endosafe nexgen-PTS)

MethodKinetic chromogenic
Lysate sensitivity0.005 EU/mL
Interference controlPositive product control recovery 50% to 200% required
DilutionSample diluted to a non-interfering concentration below the MVD

How it is reported. Reported in EU/mg of peptide. No universal endotoxin threshold exists for research-use materials, so the value is reported for the submitter's own assessment rather than judged against a limit we invented.

METALS

Elemental impurities

USP <232> limits, USP <233> procedure

ICP-MS with collision cell

ElementsCadmium, lead, arsenic, mercury (Class 1)
PreparationClosed-vessel microwave digestion in nitric acid
Internal standardsScandium, rhodium, iridium
Quantitation limit0.05 µg/g per element
Spike recovery acceptance70% to 150%

How it is reported. Reported in µg/g against the USP <232> parenteral option where the submitter designates one. Otherwise reported as measured concentrations.

MICRO

Microbial screen

USP <61> and <62> enumeration and specified organisms, with qPCR bioburden screen

Real-time PCR thermocycler (QuantStudio class)

Total aerobic microbial countUSP <61>, plate count
Total yeast and mould countUSP <61>, plate count
Specified organismsUSP <62>, absence testing
Rapid screenqPCR bioburden panel for a same-week indication

How it is reported. This is a microbial screen. It is not a USP <71> compendial sterility test, which requires a 14-day incubation and a defined aseptic sampling plan. Any certificate carrying this method says so in the method column. If you need USP <71>, ask for it by name and we will quote it as referral testing under ISO/IEC 17025 clause 6.6.

FENT

Fentanyl and analog screen

Targeted LC-MS/MS screen (MRM) with confirmation on the same run

LCMS-01, targeted MRM screen with confirmation on the same run

Screen cutoff50 ng/mL
ConfirmationA positive requires qualifier ion ratios and retention time to match a reference standard
Reporting conventionND indicates not detected at the stated cutoff

How it is reported. A not-detected result means the analyte was below the stated cutoff. It is not a statement that the substance is absent at any concentration, and the certificate says so.

STER

Sterility: rapid screen and referred compendial test

qPCR microbial DNA screen in house; USP <71> by an accredited partner laboratory

Real-time PCR thermocycler (QuantStudio class)

Rapid screenBroad-range bacterial and fungal DNA targets, reported as Detected or Not detected
Compendial testUSP <71> membrane filtration or direct inoculation, 14 days, by the partner laboratory
On the certificateThe screen is named as a qPCR screen. A referred USP <71> result names the performing laboratory, its accreditation and its report number

How it is reported. A qPCR screen detects DNA and observes no growth, so it is never reported as "no growth" or as a USP <71> result. When USP <71> is ordered, an interim certificate issues with sterility pending and the final certificate replaces it under the same accession.

TFA

Residual trifluoroacetic acid

Ion chromatography with suppressed conductivity

Ion chromatograph with conductivity detector

Reported as% w/w of the lyophilised solid
LimitsLOD and LOQ printed on every result

How it is reported. TFA is a counter-ion left by reverse-phase purification. It is reported, not judged, unless the submitter supplies a specification.

RESID

Residual solvents

Headspace GC-MS against ICH Q3C class limits

Headspace sampler with single-quadrupole GC-MS

AnalytesCommon purification and synthesis solvents, each with its own LOQ
SpecificationICH Q3C permitted daily exposure limits, converted to ppm

How it is reported. Each analyte is listed with its limit of quantitation. ND means below that limit.

CCI

Container closure integrity and solubility

Dye ingress following the approach of USP <1207>; visual solubility using USP General Notices 5.30 terms

Vacuum chamber, dye bath, visual inspection station

Closure integrityPer vial, reported as integral or not integral
SolubilityReconstitution in water at the labelled volume, described in compendial terms

How it is reported. Dye ingress is a probabilistic method: a pass means no ingress was seen under the stated challenge, not that a leak of any size is impossible.

VAR

Unit-to-unit variance

Net peptide content on additional vials, each prepared and assayed separately

HPLC-01 against a qualified reference standard

VialsThree vials beyond the one used for the main assay
Reported asEach vial's value, then the mean and the RSD across vials

How it is reported. The RSD across vials includes both fill variation and method variation. The method's own repeatability is established during validation, so the two can be told apart.

STAB

Stability program

Informed by ICH Q1A(R2) storage and timepoint design

Controlled chambers, analysis on HPLC-01 and LCMS-01

Storage conditions2 to 8 °C refrigerated, and 25 °C / 60% RH ambient challenge
Timepoints30, 60 and 90 days from accession
Assays at each pointPurity and net peptide content, against the day-zero result
Retained materialDrawn from the same retained aliquot as the original test

How it is reported. Each timepoint issues its own certificate referencing the day-zero accession, so the degradation curve is a chain of verifiable documents rather than a single summary claim.

Replicates

Every quantitative result on our certificates is the mean of replicate runs, printed with the relative standard deviation so you can see the spread rather than take the average on trust.

This matters more than it sounds. A single injection produces one number and no way to know whether that number is real. A bad injection volume, an air bubble, carryover from the previous sample or noisy peak integration all produce a plausible-looking figure. Three injections reading 97.4, 97.6 and 97.5 tell you the method is under control. Three reading 97.4, 94.1 and 98.0 tell you to investigate rather than to issue a certificate.

BasisWhat varies between runsWhat it demonstrates
Replicate injectionsThe same prepared vial is injected more than once.Instrument and integration precision. Used for purity.
Replicate preparationsThe sample is weighed and dissolved separately for each run.Precision of the whole method, including preparation. Used for net peptide content.

The certificate states which basis applies to each result and prints n and the RSD next to the value. Where an RSD exceeds the method acceptance criterion, the run is investigated and repeated. It is not averaged into something presentable.

Sampling and what a result applies to

Under ISO/IEC 17025 clause 7.3, responsibility for sampling sits with whoever performed the sampling. Samples are selected and submitted by the client, so every result applies to the unit as received. It is not evidence about any other unit of that lot, and we do not present it as such. That limitation is printed on the certificate rather than left for the reader to infer.

Where a shipment arrives in a condition that could affect the result, the deviation is recorded and the certificate states it. See sample handling and retention for the condition assessment we run on receipt.

Retained material and re-testing

Every accessioned sample is aliquoted and retained for 90 days under controlled storage. Any client may request a re-test on the retained aliquot within that window, which is the only way to settle a disagreement about a result without introducing a new sample and a new set of variables. The full policy is in sample handling and retention.

Quality control on every run

A known reference standard is analysed alongside client samples on each analytical run, and its recovery is recorded. We will publish that record, with a chart of our own accuracy over time and turnaround times computed from the same database that issues the certificates, once the laboratory is operating.

Referral testing

Work outside our scope, including USP <71> compendial sterility, is placed with a subcontracted laboratory under ISO/IEC 17025 clause 6.6. Referred work is identified as referred on the certificate, and the performing laboratory is named. We do not present another laboratory's work as our own.